Sign In | Join Free | My polstate.com |
|
Brand Name : JUGE
Model Number : AG9010Cu
Certification : IOS9001, CE
Place of Origin : Nanjing, China
MOQ : 1 Unit
Price : USD2000-7000
Packaging Details : Neutral carton
Delivery Time : 5-8 working days
Payment Terms : Online order,T/T
Supply Ability : 100 Unit/Week
Test Method : Copper Reagent Spectrophotometric Method
Measuring Range : 0~30mg/L (segmented measurement, expandable)
Detection Limit : ≤0.02mg/L
Resolution : 0.001mg/L
Accuracy : ±5%
Power Supply : AC 220±10% V
Dimensions : 520H × 370W × 265D (mm)
Weight : 40kg
Copper Online Water Quality Analyzer
>>Overview
>>Principle
The product employs the spectrophotometric method for copper detection. After mixing the water sample with a buffer agent, copper is oxidized to its higher oxidation state using a strong oxidizing agent. In the presence of a buffer solution and indicator, the oxidized copper ions react with the indicator to form a colored complex. The monitor detects this color intensity change at specific wavelengths, converts it into copper concentration values, and outputs the results. The amount of colored complex formed is proportional to the total copper concentration.
>>Technical Parameter
Test Method |
Copper Reagent Spectrophotometric Method |
Measuring Range |
0~30mg/L (segmented measurement, expandable) |
Detection Limit |
≤0.02mg/L |
Resolution |
0.001mg/L |
Accuracy |
±5% |
Repeatability |
≤±5% |
Drift (24h) |
Zero drift: ±5% Span drift: ±10% |
Measurement Time |
<30min (digestion time adjustable) |
Measurement Modes |
Adjustable time interval, fixed time point, or triggered measurement |
Calibration Modes |
Automatic calibration (1~99 days adjustable) Manual calibration based on actual water samples |
Maintenance Cycle |
|
Human-Machine Interface |
Touchscreen display and command input |
Self-Diagnosis & Protection |
Self-diagnoses operational status; retains data during anomalies/power outages Automatically clears residual reactants and resumes operation after reset/power recovery |
Data Storage |
5-year data capacity |
Input Interface |
Digital input |
Output Interfaces |
1× RS232, 1× RS485 2× 4~20mA analog outputs |
Operating Environment |
Indoor use; recommended temperature: 5~28℃, humidity ≤90% (non-condensing) |
Power Supply |
Voltage: AC 220±10% V Frequency: 50±0.5 Hz Power: ≤150W (excluding sampling pump) |
Dimensions |
520H × 370W × 265D (mm) |
![]() |
Copper Online Water Quality Analyzer Images |